Type: Control
Kit Composition:
Level 1: Human Serum, 2 x 2 mL, lyophilized
Level 2: Human Serum, 2 x 2 mL, lyophilized
Intended Use
The K-ASSAY® ASO / RF / CRP Control is intended to be used as a consistent test sample of known concentrations for monitoring the performance of the K-ASSAY® immunoturbidimetric assays for the following constituents:
Anti-Streptolysin O (ASO)
Rheumatoid Factor, RF (Ver.2)
C-Reactive Protein (CRP)
Wide Range CRP
FOR IN VITRO DIAGNOSTIC USE.
Control Preparation
Carefully and slowly remove rubber stopper and let air enter vacuum-sealed bottle.
Carefully add exactly 2 mL purified water (at room temperature) to bottle. DO NOT MIX.
Replace rubber stopper.
Let bottle sit at room temperature for 10 minutes.
Mix by slowly swirling bottle numerous times. DO NOT SHAKE or induce foaming.
Continue to mix until all contents have dissolved completely.
Before each subsequent use, be sure to mix bottle by slowly swirling.
Storage and Handling
Store lyophilized and reconstituted controls at 2–8°C. Return all controls to 2–8°C promptly after use.
Unopened controls can be used for one year from the date of manufacturing, as indicated by the expiration date on the package and bottle labels.Reconstituted controls can be used for 2 weeks if stored at 2–8°C.
Reconstituted controls may also be frozen immediately after reconstitution at –20°C for 2 months and then thawed ONE TIME ONLY.